Calibration and device tracking
The calibration plan for measuring devices, reading the result and what to do with a device that fails.
- Level:
- Practitioner
- Who it is for:
- Maintenance and calibration lead, Quality engineer
Definition
Calibration is the comparison, under stated conditions, of the value a measuring device shows with a reference of known traceability. The result gives the difference between indication and reference and the uncertainty of that difference. Comparing this result with an acceptance limit and calling it pass or fail is a separate step: verification. Adjusting the device is not calibration itself either. Device tracking means keeping each device's card, calibration and maintenance dates and results in order. Measurements made with a device that fails are re-assessed.
When does it help?
- You make product decisions based on a measuring device's result.
- Calibration dates are scattered across a spreadsheet.
- You need to show a device's calibration record in an audit.
When does it not help?
- The device is for indication only and not used in product decisions; state this clearly on the device card.
Step-by-step method
List the devices
Open a card for every measuring device used in a product or process decision.
Set the interval
Write each device's calibration interval based on how often and how critically it is used.
Write the acceptance limit
Set the device's permitted error from the tolerance of the characteristic you measure.
Verify the result
Compare the difference and uncertainty in the report with the acceptance limit; record this verification as pass or fail.
Stop it if it fails
Take the device out of use and assess the measurements made since the last passing calibration.
Example scenario
A calliper that falls outside its limit
In a quality lab, the calibration report for a calliper shows an error beyond its acceptance limit. The device is taken out of use straight away. The team lists the batches approved with that calliper since its last passing calibration and repeats the doubtful measurements with another device.
A fictional example; the people, company and numbers are not real.
Common mistakes
- Filing the report as "calibrated" without reading it.
- Not reviewing past measurements made with a device that failed.
- Leaving spare devices out of the calibration plan.
- Filing the report without writing the acceptance limit; without a limit the result cannot be verified.
Lessons
- Setting up a calibration plan
How to set the device card, the calibration interval and the acceptance limit.
1 min
- What to do with a device that fails
Stopping the device, assessing past measurements and closing the record.
1 min
How does it work in Gemy on the floor?
Device cards, calibration and maintenance schedules and work orders share one backbone. For a calibration or intermediate check, the next date is calculated from the interval. A check that leaves an unresolved "fail" takes the device out of use; once every non-conformity is resolved, that status lifts by itself. The check is entered by the person who manages devices or by the device owner.
Terms