Calibration and device tracking

The calibration plan for measuring devices, reading the result and what to do with a device that fails.

Level:
Practitioner
Who it is for:
Maintenance and calibration lead, Quality engineer

Definition

Calibration is the comparison, under stated conditions, of the value a measuring device shows with a reference of known traceability. The result gives the difference between indication and reference and the uncertainty of that difference. Comparing this result with an acceptance limit and calling it pass or fail is a separate step: verification. Adjusting the device is not calibration itself either. Device tracking means keeping each device's card, calibration and maintenance dates and results in order. Measurements made with a device that fails are re-assessed.

Calibration

When does it help?

  • You make product decisions based on a measuring device's result.
  • Calibration dates are scattered across a spreadsheet.
  • You need to show a device's calibration record in an audit.

When does it not help?

  • The device is for indication only and not used in product decisions; state this clearly on the device card.

Step-by-step method

  1. List the devices

    Open a card for every measuring device used in a product or process decision.

  2. Set the interval

    Write each device's calibration interval based on how often and how critically it is used.

  3. Write the acceptance limit

    Set the device's permitted error from the tolerance of the characteristic you measure.

  4. Verify the result

    Compare the difference and uncertainty in the report with the acceptance limit; record this verification as pass or fail.

  5. Stop it if it fails

    Take the device out of use and assess the measurements made since the last passing calibration.

Example scenario

Example scenario

A calliper that falls outside its limit

In a quality lab, the calibration report for a calliper shows an error beyond its acceptance limit. The device is taken out of use straight away. The team lists the batches approved with that calliper since its last passing calibration and repeats the doubtful measurements with another device.

A fictional example; the people, company and numbers are not real.

Common mistakes

  • Filing the report as "calibrated" without reading it.
  • Not reviewing past measurements made with a device that failed.
  • Leaving spare devices out of the calibration plan.
  • Filing the report without writing the acceptance limit; without a limit the result cannot be verified.

Lessons

How does it work in Gemy on the floor?

Device cards, calibration and maintenance schedules and work orders share one backbone. For a calibration or intermediate check, the next date is calculated from the interval. A check that leaves an unresolved "fail" takes the device out of use; once every non-conformity is resolved, that status lifts by itself. The check is entered by the person who manages devices or by the device owner.

See it on the product page

Terms

Sources

  1. International Vocabulary of Metrology (VIM3), 2.39 calibration, Joint Committee for Guides in Metrology (JCGM) / BIPM (accessed 15 November 2026)
  2. International Vocabulary of Metrology (VIM3), 2.44 verification, Joint Committee for Guides in Metrology (JCGM) / BIPM (accessed 15 November 2026)
  3. NIST Policy on Metrological Traceability, National Institute of Standards and Technology (NIST) (accessed 15 November 2026)

Cookies

Essential cookies are needed for the site to work. Analytics cookies load only if you allow them; declining is as easy as accepting. You can change your choice at any time via the footer link.

Cookie policyPrivacy notice