Calibration and device tracking

Setting up a calibration plan

How to set the device card, the calibration interval and the acceptance limit.

Level:
Introductory
Who it is for:
Maintenance and calibration lead, Quality engineer
Duration:
1 min

In this lesson

  • List what a device card should hold
  • Justify the calibration interval
  • Derive the acceptance limit from the tolerance

A calibration plan shows which device is calibrated when, by whom and against which limit. Without a plan, calibration dates are forgotten and the problem only surfaces in an audit.

The device card

Give every device a card: name, serial number, location, owner, measuring range and the measurements it is used for. The card keeps the device's history in one place.

The calibration interval

Set the interval from how often the device is used, how critical the decisions it supports are and its past results. Consider shortening the interval for a device that has fallen outside its limit before.

The acceptance limit

The acceptance limit is the largest error the device is allowed. It is set from the tolerance of the characteristic you measure; the device's error should be a small part of that tolerance. Comparing the report with this limit is a separate step after calibration: verification. Without a written limit the result cannot be verified.

  • Record the competence of the laboratory that does the calibration.
  • Check the reference traceability in the report.
  • Include spare devices in the plan.

Check yourself

There is no score; tap a question to see the answer.

What is the acceptance limit based on?

The tolerance of the characteristic measured with the device.

When would you shorten a calibration interval?

When the device has fallen outside its limit before or is used often in critical decisions.

Sources

  1. International Vocabulary of Metrology (VIM3), 2.39 calibration, Joint Committee for Guides in Metrology (JCGM) / BIPM (accessed 15 November 2026)
  2. International Vocabulary of Metrology (VIM3), 2.44 verification, Joint Committee for Guides in Metrology (JCGM) / BIPM (accessed 15 November 2026)
  3. NIST Policy on Metrological Traceability, National Institute of Standards and Technology (NIST) (accessed 15 November 2026)