Calibration and device tracking

What to do with a device that fails

Stopping the device, assessing past measurements and closing the record.

Level:
Practitioner
Who it is for:
Maintenance and calibration lead, Quality engineer
Duration:
1 min

In this lesson

  • Take a failed device out of use
  • Identify the affected measurements
  • Know the condition for returning the device to use

If the calibration report shows the device outside its acceptance limit, there are two jobs: stop the device and look back.

Stop the device

Take the device out of use straight away and put a visible mark on it. If there is a spare device, measurement continues with it.

Look back

List the measurements made with the device since its last passing calibration. Which products or batches were approved with them? Would the direction and size of the error change those decisions? Repeat doubtful measurements with another device.

Returning it to use

The device is adjusted or repaired and calibrated again. It returns to use only if the new result is within the acceptance limit. Write down the cause too, such as a drop, wear or misuse.

  • Record the date it was stopped and who stopped it.
  • Link the list of affected batches to the quality record.
  • Review the calibration interval if failures recur.

Check yourself

There is no score; tap a question to see the answer.

What are the first two jobs for a device that fails?

Take it out of use and assess the measurements made since its last passing calibration.

When does the device return to use?

After adjustment or repair, when the new calibration result is within the acceptance limit.

Sources

  1. International Vocabulary of Metrology (VIM3), 2.39 calibration, Joint Committee for Guides in Metrology (JCGM) / BIPM (accessed 15 November 2026)
  2. International Vocabulary of Metrology (VIM3), 2.44 verification, Joint Committee for Guides in Metrology (JCGM) / BIPM (accessed 15 November 2026)
  3. NIST Policy on Metrological Traceability, National Institute of Standards and Technology (NIST) (accessed 15 November 2026)

Cookies

Essential cookies are needed for the site to work. Analytics cookies load only if you allow them; declining is as easy as accepting. You can change your choice at any time via the footer link.

Cookie policyPrivacy notice